Improved Design for a Two-Piece Dental Implant System Using Fatigue and Torsion Testing Simulation

Danang Yudistiroa, Yung-Chang Chengb and Cho-Pei Jiangc
aProgram in Engineering Science and Technology, College of Engineering, National Kaohsiung University of Science and Technology, Taiwan 824, ROC.
bDepartment of Mechatronics Engineering, National Kaohsiung University of Science and Technology, Kaohsiung, Taiwan 824, ROC.
cDepartment of Mechanical Engineering, National Taipei University of Technology, Taipei, Taiwan 106, ROC


Abstract: a fatigue and torsion testing simulation and auniform design of experiment. ISO 14801 fatigue andISO 13498 torsion testing standards for the fatiguesafety factor and the von Mises stress for a two-pieceimplant model are determined usingANSYS/Workbench software. The two-piece implantmodel includes an implant and an abutment with ascrew. The six control factors for a dental implantsystem are used to increase the fatigue safety factorand decrease the von Mises stress. All control factorsare continuous in the design space so a uniformdesign is used to construct a group of simulationexperiments. A suitable uniform table is used for theuniform design to increase the strength. After theuniform design of experiment, the improved designmodel is achieved using the desirability method. Theimproved design has a maximum increase in thefatigue safety factor of 9.6 % and the von Misesstress is reduced by 6%, compared to the originaldesign. The strength of the implant is increased usinga uniform design procedure.


Keywords:  Dental implant, Fatigue safety factor, von Mises stress, ANSYS/Workbench, Uniform design, Desirability method.

Download PDF
*Corresponding author; e-mail: 
© 2022  CSME , ISSN 0257-9731 





TOP